SpectroJet | Scanning Spectrophotometer Specifications

Measurement method:
Spectral remission measurement and color density measurement according to ISO 5-3/4

Measurement geometry:
0/45° optics to DIN 5033

Measurement aperture
1,5 mm, appropriate for measuring patches with at least 3 mm height and 3 mm width. UV cut filter optional

Light source
Gas filled lamp, type A illumination

Polarization filter:
Twice linear crossed, switched on and off per software command

Measurement time:
typ. 160 mm/s; max. 400 mm/s (for color bars with 8mm width); single measurement approx. 1 second

White reference:
Absolute and relative

Illumination types / Standard observer:
A, C, D50, D65, F / 2°, 10°

Density filter:
DIN 16536, DIN 16536 NB, ISO/ANSI T, ISO/ANSI I, ISO E, spectral density Dmax

Density measurement range:
0.00 – 2.50 D

Repeatability:
0.01 D, 0,03 CIE dEab

Production Spread:
0.01 D, 0,3 CIE dEab

Data transmission:
USB-connection

Power Supply:
AC Adapter 110 – 240 V, 47 – 63 Hz

Dimensions / Weight:
70 x 135 mm, 360 grams

Accessories:
Software TECHKON ExPresso Basic
Software TECHKON ExPresso Pro
Print Control Strip TCS Digital
Horizontal Guidance Track in different lengths